Modular microfluidic system for on-chip extraction, preconcentration and detection of the cytokine biomarker IL-6 in biofluid

8 juin 2022Scientific Reports

DOI : 10.1038/s41598-022-13304-z

Auteurs

Lucile Alexandre, Amel Bendali, Iago Pereiro, Madad Azimani, Simon Dumas, Laurent Malaquin, Thanh Duc Mai, Stéphanie Descroix

Résumé

Abstract

The cytokine interleukin 6 (IL-6) is involved in the pathogenesis of different inflammatory diseases, including cancer, and its monitoring could help diagnosis, prognosis of relapse-free survival and recurrence. Here, we report an innovative microfluidic approach that uses the fluidization of magnetic beads to specifically extract, preconcentrate and fluorescently detect IL-6 directly on-chip. We assess how the physical properties of the beads can be tuned to improve assay performance by enhancing mass transport, reduce non-specific binding and multiply the detection signal threefold by transitioning between packed and fluidization states. With the integration of a full ELISA protocol in a single microfluidic chamber, we show a twofold reduction in LOD compared to conventional methods along with a large dynamic range (10 pg/mL to 2 ng/mL). We additionally demonstrate its application to IL-6 detection in undiluted serum samples.

Membres

STEPHANIE DESCROIX

Directeur de recherche CNRS